Field flow fractionation

Field-flow fractionation, abbreviated FFF, is a separation technique in which a field (thermal, electric, magnetic, hydraulic, gravitational, ...) is applied to a diluted suspension in a fluid or to a solution pumped through a long and narrow channel, perpendicular to the direction of the field, in order to cause the separation of particles present in the fluid, depending on their differing "mobilities" under the force exerted by the field. It was devised and reported by J. Calvin Giddings in 1976.[1] Giddings, credited for the invention of FFF, was professor of chemistry and specialist of chromatography and separation techniques at the University of Utah. The FFF method is unique to other separation techniques because it can separate materials over a wide colloidal size range while maintaining high resolution. Although FFF is an extremely versatile technique, there is no "one size fits all" method for all applications.

Flow field-flow fractionation (AF4) channel cross section, where the velocity of laminar flow within the channel is not uniform. The fluid travels in a parabolic pattern with the velocity of the flow, increasing towards the centre of the channel and decreasing towards the sides.

In field-flow fractionation the field can be hydraulical (with an asymmetrical flow through a semi-permeable membrane), gravitational, centrifugal, thermal, electrical, or magnetic. In all cases, the separation mechanism is produced by differences in particle mobility (electrophoretic, when the field is a direct current (DC) electric field causing a transverse electric current flow) under the forces of the field, in equilibrium with the forces of diffusion: an often-parabolic laminar-flow-velocity profile in the channel determines the velocity of a particular particle, based on its equilibrium position from the wall of the channel. The ratio of the velocity of a species of particle to the average velocity of the fluid is called the retention ratio.

Fundamental principles

Field flow fractionation is based on laminar flow of particles in a solution. These sample components will change levels and speed based on their size/mass. Since these components will be travelling at different speeds, separation occurs. A simplified explanation of the setup is as follows. The sample separation occurs in a thin, ribbon-like, channel in which there is an inlet flow and a perpendicular field flow. The inlet flow is where the carrier liquid is pumped into the channel and it creates a parabolic flow profile and it propels the sample towards the outlet of the channel. Once the sample elutes from the channel it enters the detector. Due to FFF's similarity to Liquid Chromatography, in ways of a liquid mobile phase passing through the channel, the most common detectors are those that are also used for LC. The most frequently used is a UV detector because of its non-destructive nature.

Relating force (F) to retention time (TR)

The relationship between the separative force field and retention time can be illustrated from first principles. Consider two particle populations within the FFF channel. The cross field drives both particle clouds towards the bottom "accumulation" wall. Opposing this force field is the particles natural diffusion, or Brownian motion, which produces a counter acting motion. When these two transport process reach equilibrium the particle concentration c approaches the exponential function of elevation x above the accumulation wall as illustrated in equation 1.

l represents the characteristic elevation of the particle cloud. This relates to the height that the particle group can reach within the channel and only when the value for l is different for either group will separation occur. The l of each component can be related to the force applied on each individual particle.

Where k is the Boltzmann constant, T is absolute temperature and F is the force exerted on a single particle by the cross flow. This shows how the characteristic elevation value is inversely dependent to the Force applied. Therefore, F governs the separation process. Hence, by varying the field strength the separation can be controlled to achieve optimal levels. The velocity V of a cloud of molecules is simply the average velocity of an exponential distribution embedded in a parabolic flow profile. Retention time, tr can be written as:

Where L is the channel length. Subsequently, the retention time can be written as:

tr/to = w/6l ⌊coth w/2l- 2l/w⌋−1

Where to is the void time (emergence of a non-retained tracer) and w is the sample thickness. Substituting in kT/F in place of l illustrates the retention time with respect to the cross force applied.

tr/to = Fw/6kT ⌊coth Fw/2kT- 2kT/Fw⌋−1

For an efficient operation the channel thickness value w far exceeds l. When this is the case the term in the brackets approaches unity. Therefore, equation 5 can be approximated as:

tr/to = w/6l = Fw/6kT

Thus tr is roughly proportional to F. The separation of particle bands X and Y, represented by the finite increment ∆tr in their retention times, is achieved only if the force increment ∆F between them is sufficient. A differential in force of only 10–16 N is required for this to be the case. The magnitude of F and ∆F depend on particle properties, field strength and the type of field. This allows for variations and specialisations for the technique. From this basic principle many forms of FFF have evolved varying by the nature of the separative force applied and the range in molecule size to which they are targeted.

Fractogram

Centrifugal FFF separates by mass (i.e. a combination of particle density and particle size). For example, gold and silver nanoparticles of identical size can be separated into two peaks, according to differences in density of gold and silver.

A graph of detection signal vs. time from an FFF separation, in which various substances present in a fluid are separated based on their flow velocities under some applied external field, such as an flow, centrifugal, thermal or electric field.

Often these substances are particles initially suspended in a small volume of a liquid buffer and pushed along the FFF channel by the buffer. The varying velocities of a particular species of particles may be due to its size, its mass, and/or its distance from the walls of a channel with non-uniform flow-velocity. The presence of different species in a sample can thus be identified through detection of a common property at some distance down the long channel, and by the resulting fractogram indicating the presence of the various species by peaks, due to the different times of arrival characteristic of each species and its physical and chemical properties.

In an electrical FFF, an electric field controls the velocity by controlling the lateral position of either a charged (having electrophoretic mobility) or polarized (being levitated in a non-uniform field) species in a capillary channel with a hydrodynamically parabolic flow-velocity profile, meaning that the velocity of the pumped fluid is highest midway between the walls of the channel and it monotonically decays to a minimum of zero at the wall surface.[2]

Forms

Most techniques available today are advances on those originally created by Prof. Giddings nearly 4 decades ago.

Flow

Of these techniques flow FFF was the first to be offered commercially. Flow FFF separates particles based on size, independent of density and can measure macromolecules in the range of 1 nm to 1 µm. In this respect it is the most versatile FFF sub-technique available. The cross flow in Flow FFF enters through a porous frit at the top of the channel, exiting through a semi-permeable membrane outlet frit on the accumulation wall (i.e. the bottom wall).

Hollow fiber flow

Hollow fiber flow FFF (HF5) was developed by Lee et al. (1974).[3] HF5 has been applied towards the analysis of lattices and other macromolecules. HF5 was the first form of flow FFF to be developed in 1974. Flat membranes soon outperformed hollow fibers and forced HF5 into obscurity. One of the drawbacks of HF5 is the availability of the membranes with uniform pore sizes. There are different kinds of ceramic and polymeric hollow fiber membranes used in practice.

Asymmetric flow

Asymmetric flow FFF (AF4), on the other hand, has only one semi-permeable membrane on the bottom wall of the channel. The cross flow is, therefore, created by the carrier liquid exiting the bottom of the channel. This offers an extremely gentle separation and an “ultra-broad” separation range. High Temperature Asymmetric Flow Field-Flow Fractionation is the most advanced technology for the separation of high and ultra-high molar mass polymers, macromolecules and nanoparticles in the size range.

Thermal

Thermal FFF, as the name suggests, establishes a separation force by applying a temperature gradient to the channel. The top channel wall is heated and the bottom wall is cooled driving polymers and particles towards the cold wall by thermal diffusion. Thermal FFF was developed as a technique for separating synthetic polymers in organic solvents. Thermal FFF is unique amongst FFF techniques in that it can separate macromolecules by both molar mass and chemical composition, allowing for the separation of polymer fractions with the same molecular weight. Today this technique is ideally suited for the characterization of polymers, gels and nanoparticles.

One of the major advantage of Thermal FFF is the simple and very well defined dimensions of the separation channel, which makes the inter-lab or inter-instrument Universal Calibration possible because the Thermal FFF calibration constants closely describe the ratio of ordinary (molecular) diffusion coefficient D to thermal diffusion coefficient (or, thermophoretic mobility) DT which are only polymer dependent. The ThFFF Universal Calibration is, therefore, instrument and lab transferable, while the well-known size exclusion chromatography Universal Calibration is polymer-transferable on the same instrument only. [4]

Split flow thin-cell fractionation

Split flow thin-cell fractionation (SPLITT) is a special preparative FFF technique, using gravity for separation of µm-sized particles on a continuous basis. SPLITT is performed by pumping the sample containing liquid into the top inlet at the start of the channel, whilst simultaneously pumping a carrier liquid into the bottom inlet. By controlling the flow rate ratios of the two inlet streams and two outlet streams, the separation can be controlled and the sample separated into two distinct sized fractions. The use of gravity alone as the separating force makes SPLITT the least sensitive FFF technique, limited to particles above 1 µm.

Centrifugal

In centrifugal FFF, the separation field is generated via a centrifugal force. The channel takes the form of a ring, which spins at speeds up to 4900 rpm in the case of Postnova Analytics CF2000 instrument. The flow and sample are pumped into the channel and centrifuged, allowing the operator to resolve the particles by mass (size and density). The advantage of centrifugal FFF lies in the high size resolution that can be achieved by varying the force applied, since particle size is proportional to particle mass to the third power.

The unique advantage presented by centrifugal FFF comes from the techniques capability for high resolution. The only commercially available centrifugal FFF instrument is Postnova Analytics' CF2000, incorporating the unique feature of separating particles by both particle size and density. This allows for the separation of particles with only a 5% difference in size.

Centrifugal FFF has the advantage that particles and macromolecules can be separated by particle density, rather than just particle size. In this instance, two identically sized gold and silver nanoparticles can be separated into two peaks, according to differences in density in the gold and silver nanoparticles, separated with the Centrifugal FFF Postnova CF2000 instrument with detection by Dynamic Light Scattering (DLS).

In AF4 separations, the ratio of mass to time is 1:1. With the addition of the third parameter of density to centrifugal FFF, this produces a ratio more akin to mass:time to the power of three. This results in a significantly larger distinction between peaks and result in a greatly improved resolution. This can be particularly useful for novel products, such as composite materials and coated polymers containing nanoparticles, i.e. particles which may not vary in size but which do vary in density. In this way two identically sized particles can still be separated into two peaks, providing that the density is different.

References

  1. Giddings, J.C., Yang F.J., and Myers M.N. (1976). "Flow Field-Flow Fractionation: a versatile new separation method.” Science 193.4259: 1244–1245.
  2. Madou, Marc (2001). Fundamentals of Microfabrication. US: CRC. pp. 565–571. ISBN 0-8493-0826-7.
  3. Lee H.L., Reis J.F.G., and Lightfoot E.N. (1974). Single-phase chromatography: Solute retardation by ultrafiltration and electrophoresis. AIChE Journal, vol. 20, p. 776.
  4. W.J. Cao, P.S. Williams, M. N. Myers, and J.C. Giddings, “Thermal Field-Flow Fractionation Universal Calibration: Extension for Consideration of Variation of Cold Wall Temperature”, Analytical Chemistry, 1999, 71, pp1597 – 1609
This article is issued from Wikipedia. The text is licensed under Creative Commons - Attribution - Sharealike. Additional terms may apply for the media files.